014052 : Screw type reference electrode for alkaline solution
The RE-61VT reference electrode is made of Hg/HgO instead of calomel paste, and 1 M sodium hydroxide is used as an electrolyte solution. It has the specific design to be suitable for the ALS line-up flow cell, as Cross flow cell, Radial flow cell, EQCMT flow cell and SEC-3F Spectroelectrochemical flow cell. Also, potassium hydroxide solution can be used as a internal solution, instead of sodium hydroxide solution.
Reference potential |
HgO + H2O + 2e = Hg + 2OH- E0=118 mV vs RHE (25 deg C) ∗ The reference potential is the measured value using RHEK reversible hydrogen electrode kit (Cat. No. 013597). |
This product is designed to be filled with 1M NaOH solution by the user. Please handle with care. ∗Please note that we do not sell the 1M NaOH solution to be used.
Holder material: Polymethylpentene, for the high chemical resistance is used.
Main body: Screw type to be removable
Liquid junction: Ceramic tip
From delivery to use :
This reference electrode is designed for users to fill the electrode body with 1M NaOH solution themselves after receiving it. We recommend leaving it for at least 3 hours between filling and use to stabilize the potential.
Note : When using with Cross flow cell, Radial flow cell, EQCMT flow cell and SEC-3F, please avoid the below conditions, because some contents and working electrodes have a low tolerance for alkaline solution. ➀ High concentrated alkaline solution ➁ High temperature condition ➂ Long time application
∗Keep the reference electrode in a cold dark place.
Quantity
Catalog
Catalog No.
|
Description
|
Electrolyte
|
Purpose
|
014052
|
RE-6VT Reference electrode screw type
|
-
|
Reference electrode for alkaline solution
|