top of page

BVT SPE with immobilized Acetylcholinesterase

Amperometric Acetylcholinesterase three-electrode sensor

  • Based on AC1.W2.RS
  • WE has immobilized acetylcholinesterase
  • Detect pesticides, toxic and neurotoxic gases

Quantity

Description

Detect pesticides with this enzyme modified electrode.

Acetylcholinesterase from electric eel type VI-S is immobilized on the working electrode’s active surface of an electrochemical sensor AC1.W2.RS (i.e. Platinum working electrode, silver reference electrode). The diameter of the immobilized bioactive membrane is 2 mm and the mean applied activity is 1 unit/mm².

What can the sensor detect?

With this electrode you can detect acetylcholinesterase inhibitors by measuring the enzyme activity before and after exposure to your sample. Analytes like organophosphorous and carbamate pesticides, toxic and neurotoxic gases (e.g. sarin) and chemical weapons inhibit acetylcholinesterase and can thus be detected.

Dimensions

  • 25.4 x 7.26 x 0.63 mm

More information

  • A detailed description of these sensors can be found in the literature: Analyst, 2002, 127, 5–7

Ref. BVT-AC1.AChE

Specifications

WE material : Platinum

Surface modification : Acetyl Cholinesterase

Brand : BVT Technologies

Amount of WEs : 1

CE material : Platinum

Electrode system : CE + RE + WE

Combined RE and CE : No

Ø of the disc WE : 1 mm

WE geometric area : 0.79 mm²

RE material : Silver

Support material : Alumina Ceramic

Support dimensions : 7 by 25 mm

Possible analytes : chemical weapons by measuring the enzyme activity before and after immersion in the sample, organophosphorous and carbamate pesticides, Pesticides, Acetylcholinesterase inhibitors, toxic and neurotoxic gases (e.g. sarin)

Sensor type : Screen-printed

bottom of page