top of page

CE-6002-1500V200A

1500V Ultra-High Voltage Battery Cluster Solution

Quantity

Key Highlights

  • Ultra-high-voltage testing up to 1500 V

  • Current output up to 200 A per channel

  • 300 kW single-channel output power

  • 600 kW total output power

  • Voltage & current accuracy: ±0.02% F.S.

  • 24-bit voltage and current resolution

  • Current response time ≤10 ms

  • Current conversion time ≤20 ms

  • Minimum recording interval: 10 ms

  • Parallel operation of up to 4 channels

  • AC/DC + DC/DC dual-stage high-frequency isolated architecture

  • TCP/IP, CAN FD, CAN, and RS485 communication

  • BMS communication and integration

  • Built-in DCIR testing

  • Energy feedback efficiency up to 96%

  • Designed for energy storage packs, high-voltage battery clusters, and PCS testing

Description

The NEWARE CE-6002-1500V200A is an ultra-high-voltage battery testing system designed for energy storage packs, high-voltage battery clusters, and high-voltage DC source simulation for PCS testing. Featuring a voltage output range of up to 1500 V, current output of up to 200 A, and 300 kW of single-channel power, the system provides high-power testing capabilities for next-generation energy storage and high-voltage battery applications. Its AC/DC and DC/DC dual-stage high-frequency isolated modular architecture provides a scalable and flexible platform for R&D and production testing. Up to four channels can be operated in parallel to support higher-current applications, with parallel expansion available up to 1,600 A depending on configuration. The system also supports TCP/IP Ethernet, CAN FD, CAN, RS485, and BMS communication, enabling integration with automated testing environments and peripheral equipment such as temperature chambers, water chillers, and vibration testing systems.

Specifications

Parameter

Specification

Equipment Name

1500V Ultra-High Voltage Battery Cluster Solution

Model

CE-6002-1500V200A

Voltage Output – Charge

0–1500 V

Voltage Output – Discharge

100–1500 V

Minimum Discharge Voltage

100 V

Voltage Accuracy

±0.02% F.S.

Voltage Resolution

24-bit

Current Output Range

1–200 A

Current Accuracy

0.02% F.S. per independent range

CV Cut-off Current

0.05% F.S. per independent range

Current Resolution

24-bit

Single-Channel Output Power

300 kW

Total Output Power

600 kW

Current Response Time

≤ 10 ms

Current Conversion Time

≤ 20 ms

Minimum Recording Interval

10 ms

Parallel Operation

Up to 4 channels

Parallel Expansion

Up to 1,600 A, depending on configuration

Communication

TCP/IP Ethernet, CAN FD, CAN, RS485

BMS Integration

Supported

DCIR Testing

Built-in

Energy Feedback Efficiency

Up to 96%

Architecture

AC/DC + DC/DC dual-stage high-frequency isolated

The core technical specifications above are taken from NEWARE's current product page.

Application

Energy Storage Pack Testing

High-voltage charge/discharge testing and performance evaluation of energy storage battery packs.

High-Voltage Battery Cluster Testing

Suitable for testing high-voltage battery clusters used in large-scale energy storage and power applications.

PCS Testing

High-voltage DC source simulation for Power Conversion System (PCS) testing.

Battery Performance Evaluation

Supports charge/discharge characterization, DCIR testing, and high-power battery performance evaluation.

BMS Testing

Integration with mainstream BMS communication systems for battery management and system-level testing.

Environmental & Condition Simulation

Can be integrated with environmental chambers, water chillers, vibration testers, and other peripheral equipment for comprehensive battery testing.

bottom of page