CT-4008Q-5V12A-S1
Highly accurate 4-range tester, master of detail realism. The NEWARE battery tester is engineered with integrated probes specifically for cylindrical batteries, allowing for height adjustment to accommodate various sizes of cylindrical batteries in testing.
Quantity
Key Highlights
8 independent channels with independent data acquisition and programming
Voltage range up to 5 V
Current range from 0.5 mA to 12 A
4 selectable current ranges
Current and voltage accuracy of ±0.02% F.S.
Current and voltage stability of ±0.02% F.S.
60 W output power per channel
10 Hz recording frequency
100 ms sampling time
Current response time ≤1.5 ms in the technical specifications
500 ms minimum pulse width
Up to 254 programmable work-step lines
1 GB offline testing
Supports CC/CV charge and discharge
Supports GITT, DCIR, pulse simulation, cycle life, and rate charge/discharge testing
Supports dQ/dV differential capacity analysis
3U modular design
Integrated adjustable probes designed for cylindrical batteries
Supports auxiliary voltage, auxiliary temperature, and environmental test chamber integration
Description
The NEWARE CT-4008Q-5V12A-S1 is an 8-channel battery testing system designed for precise charge/discharge characterization of rechargeable batteries and energy storage devices. Each channel can independently perform data acquisition and be programmed according to individual testing requirements. The system provides four selectable current ranges from 0.5 mA to 12 A: 0.5–100 mA, 100–500 mA, 500 mA–6 A, and 6–12 A. This range switching allows users to select an appropriate current range for improved measurement accuracy, including low-current testing. With 60 W output power per channel, the CT-4008Q-5V12A-S1 is designed for higher-current battery testing than the 6 A version. It supports a range of testing methods, including cycle life, rate charge/discharge, pulse simulation, GITT, DCIR, and dQ/dV analysis. The system also supports DBC protocol configuration, CC/CV charge-discharge testing, segmented data recording, and connection to auxiliary temperature, voltage, and environmental test equipment. Its 3U modular design allows multiple units to be assembled into larger battery testing systems.
Specifications
Parameter | Specification |
Equipment Name | CT-4000 Battery Testing System |
Model | CT-4008Q-5V12A-S1 |
Number of Channels | 8 channels |
Channel Operation | Independent data acquisition & programming |
CV Output Range | 25 mV – 5 V |
Minimum Discharge Voltage | 2 V with universal holder; 2.5 V with 2 m cable |
Voltage Accuracy | ±0.02% F.S. |
Voltage Stability | ±0.02% F.S. |
Current Range 1 | 0.5 mA – 100 mA |
Current Range 2 | 100 mA – 500 mA |
Current Range 3 | 500 mA – 6 A |
Current Range 4 | 6 A – 12 A |
Current Accuracy | ±0.02% F.S. |
Current Stability | ±0.02% F.S. |
CV Cut-off Current | 0.2 mA / 1 mA / 12 mA / 24 mA |
Output Power | 60 W/channel |
Power Stability | ±0.04% F.S. |
Recording Frequency | 10 Hz |
Sampling Time | 100 ms |
Current Response Time | ≤1.5 ms (10–90% F.S.) |
Minimum Pulse Width | 500 ms |
Offline Testing | 1 GB |
Programmable Work Steps | Up to 254 editable lines |
Working Step Time | ≤365 × 24 h/step |
Design | 3U modular |
F.S. = Full Scale. Specifications follow NEWARE's current product page. The page's feature section states a response time of ≤1 ms, while the detailed technical specification states ≤1.5 ms; for the formal RYTS specification table, I recommend using the detailed value of ≤1.5 ms.
Application
The CT-4008Q-5V12A-S1 is suitable for:
Rechargeable Battery Testing
Battery Cell Characterization
3C Electronics Battery Testing
Consumer Electronics
Communication Devices
Laptop & Tablet Battery Testing
Battery Materials Research
Solid-State Battery Research
Cycle Life Testing
Rate Charge / Discharge Testing
GITT (Galvanostatic Intermittent Titration Technique)
DCIR Testing
Pulse Simulation
dQ/dV Differential Capacity Analysis
NEWARE specifically lists Consumer Electronics, Communication, and Laptops/Tablets for this model, while the product supports broader battery research methods, including GITT, DCIR, pulse simulation, cycle life, and dQ/dV analysis.