Reference electrode for Non Aqueous solution

  • For apply in an organic solvent sample
  • Internal solution changeable

Preservation of the electrode

  • Fill the sample holder with an acetonitrile/0.1 M tetrabutylammonium perchlorate/0.01 M AgNO3 solution.
  • Rinse the Ag wire with acetone.
  • Keep the electrode immersed in an acetonitrile/0.1 M tetrabutylammonium perchlorate solution, with or without Ag ion (0.01 M AgNO3).

Note :

  • Avoid the dry of the IPPG (Ion Permeability Porous Glass); glass, it could cause the crash of the IPPG; glass tip, caused by the salt crystallization inside of the glass. Make sure to keep the tip always wet.
  • Before assembly the electrode, immerse and keep the IPPG; glass tip in an organic solvent for at least one hour.
  • If bubbles are present around the IPPG; glass, remove with a slight tapping. Air bubbles will isolate, blocking the current flow.
  • Use only high grade products and reagents.
  • Wash with acetonitrile. Salt crystallization, in the inner side of the Teflon cap and at the edge of the sample holder, makes difficult the assembly.

Catalog

Quantity

Precautions

Electrode

This electrode is silver-silver ion reference electrode for non-aqueous solution.
In order to reduce the liquid junction potential, use the same solvent and supporting electrolyte, as a sample and electrode internal solution.

Solvent

The solvent in non-aqueous electrochemical should be considered as following :

  • Electrode activity material and produced substance solubility.
  • Select according to the potential area to be measured.

For example, dimethyl sulfoxide (DMSO) is major solvent in the negative potential area, and acetonitrile (ACN) is major solvent in the positive potential area.
The non-aqueous solution could contains impurities, and the impurities could cause the block reactions. Distilled the solvent sufficiently, to eliminated impurities before using.

Supporting electrolyte

If the sample is dissolved in an organic solvent, the supporting electrolyte must be added.
In order to select the supporting electrolyte, it is necessary to consider the following :

  • Solubility in organic solvents
  • Wide potential window
  • No reaction with organic solvent

Typical supporting electrolyte :

  • TEAP : Tetraethylammonium perchlorate
  • TBAPF6 : Tetrabutylammonium hexafluorophosphate

Assembly

Contents for, Non Aqueous reference electrode

  1. Teflon cap with Ag wire (1 pc)
  2. Sample holder dia 6mm (1 pc)

The internal solution is not included, please prepare the internal solution first. In the section Internal solution, there is an instruction for the preparation of an acetonitrile (ACN) solution containing 0.01 M silver(I) nitrate (AgNO3) and 0.1 M tetrabutylammonium perchlorate (TBAP), it can be used as a reference.​​​​​​​

  1. Immerse the sample holder tip (IIPG) in an organic solvent to be used,at least for one hour.
  2. Fill the sample holder with 0.6 mL of the electrolyte solution, use the pipette and fill it carefully.
  3. If there is a presence of the air bubble on the extremity (liquid junction), flick the sample holder with the fingers.
  4. Set the Teflon cap and seal the cap with a parafilm, to avoid the evaporation of the inner solution.
    If the reagents will be prepared by yourself, choose a special grade reagents.)

​​​​​​​Attention for the re-assembly​​​​​​​

  • There is a possibility to have the precipitation of the support salt inside of the sample holder and Teflon cap, please wash with acetonitrile before re-use.
  • If the extremity tip (liquid junction IPPG) is dried, suport salt deposition will happen, blocking the holes which increases the liquid junction resistance or breaks the tip (IPPG). Keep it moistening in a solvent.
    Ion Permeability Porous Glass clogging can not be recovered.

Catalog

  • Catalog No.

Description

Junction

Purpose

013848

RE-7N Non Aqueous reference electrode

IPPG

SVC-2, SVC-3, VC-4, SBC Bulk electrolysis, RRDE, QCMT

013849

RE-7SN Non Aqueous reference electrode

IPPG

SECM

013850

RE-7VN Non Aqueous reference electrode screw type

Ceramics

Flow cell (Cross, Radial, EQCMT, SEC-3F)

Optional items (sold separately)

012108

RE-PV Preservative vial for reference electrode

For preservation of OD 6.0 mm reference electrode

012057

RE-7 Teflon cap with Ag wire

With Ag wire

012058

RE-7S Teflon cap with Ag wire

With Ag wire

012176

Sample holder dia 6 mm (2 pcs)

IPPG

For double junction, manufacturing of the electrode

012177

Sample holder dia 9 mm (2 pcs)

IPPG

For double junction, manufacturing of the electrode